Oscilloscope to PC Connection Software
Select Model for Instructions
Model 54621A, Model 54603B & 54645D.
54603B 2-Channel 60
MHz Oscilloscope
54645D 2-Channel 100-MHz 200-MSa/s Mixed Signal Oscilloscope
Set Oscilloscope Baud Rate Setting
The oscilloscope Baud Rate must match the IntuiLink Baud Rate setting.
Select oscilloscope Utility button then select I/O.
Set Baud Rate to 19200 and Handshake to DTR as displayed below.

IntuiLink 54603B & 54645D Setup
Agilent IntuiLink Data Capture Program is the connectivity software that provides convenient access to instruments within Microsoft Windows.
To run IntuiLink see the start menu example below.
Welcome Box
If the Welcome Box opens: √ Check Mark "Do not show this dialog again" and close the box.

Select 54600-Series
Get Waveform Data Tab
IntuiLink is capable of downloading both the oscilloscope Screen Image and the Waveform Data (useful for spreadsheets and data analysis).
Uncheck mark field Get waveform data to download only the Screen Image.
Set I/O Tab
Select the Set I/O Tab then Find Instruments button
Identify Instrument(s)
Select Identify Instrument(s) button
IntuiLink COM1 Settings
Set IntuiLink Baud Rate to 19200
The instrument found should be displayed as exampled below. If your instrument shows up as ??? question marks try turning off/on your oscilloscope and try again.
Select the Connect button
Instrument should be displayed as Connected. Select Close button.
Select the OK button
Oscilloscope will be frozen for a minute as the download of the screen image occurs.
When completed oscilloscope screen will be displayed within IntuiLink.